Temperature test chamber 335 litersVötsch
VT 4034 -40°C..+180°C
Temperature test chamber 335 liters
Vötsch
VT 4034 -40°C..+180°C
وضعیت
کارکرده
موقعیت
Borken 

تصاویر نشان میدهند
نمایش نقشه
اطلاعات دستگاه
قیمت و مکان
- موقعیت:
- Einsteinstraße 8a, 46325 Borken, DE
تماس بگیرید
جزئیات پیشنهاد
- شناسه آگهی:
- A18998656
- شماره مرجع:
- 24602
- بهروزرسانی:
- آخرین بهروزرسانی در ۰۳.۰۲.۱۴۰۴
توضیحات
Vötsch
Type: VT 4034
Vötsch Temperature Cabinet VT 4034 Air-Cooled
Refrigerant: Converted to R-452 A
Internal dimensions: Height 750 mm, Depth 765 mm, Width 580 mm
External dimensions: Height 1780 mm, Depth 1595 mm, Width 865 mm
Temperature range: -40°C to +180°C
Temperature deviation:
temporal ± 0.1 to ± 0.5 K
spatial ± 0.5 to ± 2.0 K
Heating 3.2 K/min
Cooling 3.5 K/min
Maximum load (evenly distributed across the surface)
of the test chamber floor: 60 kg
per insert: 30 kg
Total load of the grate supports: 80 kg
Test chamber lighting: 24 V halogen bulb 50W
Rated voltage: 3/N/PE AC 400V ± 10% 50 Hz
Rated power: 4.1 kW
Rated current: 12.5 A
Guedpfx Aewd Hpajggeh
Fuse provided by the customer: 16 A slow-blow
Refrigerant: R-452A
Weight: 490 kg / 335 liters
For your safety, the following information!
The following steps are carried out on our chambers in advance:
1. Functional test and replacement of necessary components
2. If necessary, refill with legally compliant refrigerant
3. Leak test with certificate
4. After successful testing, the chambers are subjected to a documented test run.
Condition: Used
Scope of delivery: (See image)
(Subject to changes and errors in the technical data, information is subject to change!)
We are happy to answer any further questions you may have over the phone.
آگهی به صورت خودکار ترجمه شده است. احتمال بروز اشتباه در ترجمه وجود دارد.
Type: VT 4034
Vötsch Temperature Cabinet VT 4034 Air-Cooled
Refrigerant: Converted to R-452 A
Internal dimensions: Height 750 mm, Depth 765 mm, Width 580 mm
External dimensions: Height 1780 mm, Depth 1595 mm, Width 865 mm
Temperature range: -40°C to +180°C
Temperature deviation:
temporal ± 0.1 to ± 0.5 K
spatial ± 0.5 to ± 2.0 K
Heating 3.2 K/min
Cooling 3.5 K/min
Maximum load (evenly distributed across the surface)
of the test chamber floor: 60 kg
per insert: 30 kg
Total load of the grate supports: 80 kg
Test chamber lighting: 24 V halogen bulb 50W
Rated voltage: 3/N/PE AC 400V ± 10% 50 Hz
Rated power: 4.1 kW
Rated current: 12.5 A
Guedpfx Aewd Hpajggeh
Fuse provided by the customer: 16 A slow-blow
Refrigerant: R-452A
Weight: 490 kg / 335 liters
For your safety, the following information!
The following steps are carried out on our chambers in advance:
1. Functional test and replacement of necessary components
2. If necessary, refill with legally compliant refrigerant
3. Leak test with certificate
4. After successful testing, the chambers are subjected to a documented test run.
Condition: Used
Scope of delivery: (See image)
(Subject to changes and errors in the technical data, information is subject to change!)
We are happy to answer any further questions you may have over the phone.
آگهی به صورت خودکار ترجمه شده است. احتمال بروز اشتباه در ترجمه وجود دارد.
تأمینکننده
توجه: به صورت رایگان ثبتنام کنید یا وارد شوید, تا به تمام اطلاعات دسترسی داشته باشید.
ارسال درخواست
تلفن & فکس
+49 2861 ... آگهیها