Climatic chamber with humidity, 34 litersWeiss
WKL 34/60 -60°C bis 150°C
Climatic chamber with humidity, 34 liters
Weiss
WKL 34/60 -60°C bis 150°C
وضعیت
کارکرده
موقعیت
Borken 

اطلاعات دستگاه
قیمت و مکان
- موقعیت:
- Einsteinstraße 8a, 46325 Borken, Deutschland

تماس بگیرید
جزئیات پیشنهاد
- شناسه آگهی:
- A22199292
- شماره مرجع:
- 24134
- بهروزرسانی:
- آخرین بهروزرسانی در ۰۳.۰۴.۱۴۰۵
توضیحات
Weiss
Type: WKL 34/60
Weiss Climate Chamber, Air-Cooled
Refrigerant converted to R-452A
-60°C to 150°C
Volume: 34 liters
Installed in basic configuration:
R1 NW 50 mm
L1NW 120 mm
Internal dimensions: Height 310 mm, Width 324 mm, Depth 300 mm
External dimensions: Height 1740 mm, Width 630 mm, Depth 970 mm
Performance data for temperature tests:
Temperature range -60 to +150 °C
Temperature differences in steady state, according to IEC 60068-3-5
Temperature fluctuation over time in the center of the usable space ± 0.3 to ± 1 K ± 1 to ± 2 K
Temperature deviation spatially (corresponds to temperature gradient) 2 to 4
Temperature change rate (according to IEC 60068-3-5)
Heating 2.5 K/min
Cooling 3 K/min
Heat compensation max. 550 W
Performance data for climate tests:
Temperature range +10 to +95 °C
Humidity range 10 to 98 % RH
Temperature and humidity differences (in steady state, according to IEC 60068-3-5)
Temperature fluctuation over time in the center of the usable space ± 0.3 to ± 0.5 K
Temperature deviation spatially ± 0.5 to ± 1.5
Lodezgr U Rspfx Acbern
(corresponds to temperature gradient) 1 to 3
Humidity fluctuation over time in the center of the usable space ± 3 to ± 5 % RH
Operating data:
Test chamber lighting Halogen bulb 12V, 20W
Rated voltage 1/N / PE AC 230 V ± 10 % 50 Hz
Rated power 2.5 kW
Rated current 11 A
Fuse on customer side 16A slow-blow
Heat dissipation of air-cooled test chambers
Max. heat dissipation to the environment 1700 W
Refrigerant: R-452A / R23
Weight: 140 kg
For your safety as a buyer, please note the following:
The following points are carried out on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. If necessary, refill with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers are subjected to a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in technical data are subject to change!)
We will gladly answer any further questions you may have over the phone.
آگهی به صورت خودکار ترجمه شده است. احتمال بروز اشتباه در ترجمه وجود دارد.
Type: WKL 34/60
Weiss Climate Chamber, Air-Cooled
Refrigerant converted to R-452A
-60°C to 150°C
Volume: 34 liters
Installed in basic configuration:
R1 NW 50 mm
L1NW 120 mm
Internal dimensions: Height 310 mm, Width 324 mm, Depth 300 mm
External dimensions: Height 1740 mm, Width 630 mm, Depth 970 mm
Performance data for temperature tests:
Temperature range -60 to +150 °C
Temperature differences in steady state, according to IEC 60068-3-5
Temperature fluctuation over time in the center of the usable space ± 0.3 to ± 1 K ± 1 to ± 2 K
Temperature deviation spatially (corresponds to temperature gradient) 2 to 4
Temperature change rate (according to IEC 60068-3-5)
Heating 2.5 K/min
Cooling 3 K/min
Heat compensation max. 550 W
Performance data for climate tests:
Temperature range +10 to +95 °C
Humidity range 10 to 98 % RH
Temperature and humidity differences (in steady state, according to IEC 60068-3-5)
Temperature fluctuation over time in the center of the usable space ± 0.3 to ± 0.5 K
Temperature deviation spatially ± 0.5 to ± 1.5
Lodezgr U Rspfx Acbern
(corresponds to temperature gradient) 1 to 3
Humidity fluctuation over time in the center of the usable space ± 3 to ± 5 % RH
Operating data:
Test chamber lighting Halogen bulb 12V, 20W
Rated voltage 1/N / PE AC 230 V ± 10 % 50 Hz
Rated power 2.5 kW
Rated current 11 A
Fuse on customer side 16A slow-blow
Heat dissipation of air-cooled test chambers
Max. heat dissipation to the environment 1700 W
Refrigerant: R-452A / R23
Weight: 140 kg
For your safety as a buyer, please note the following:
The following points are carried out on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. If necessary, refill with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers are subjected to a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in technical data are subject to change!)
We will gladly answer any further questions you may have over the phone.
آگهی به صورت خودکار ترجمه شده است. احتمال بروز اشتباه در ترجمه وجود دارد.
ارسال درخواست
تلفن & فکس
+49 2861 ... آگهیها
آگهی شما با موفقیت حذف شد
یک خطا رخ داده است